KFUPM ePrints

Measurement of light output of NE213 and NE102A detectors for2.7-14.5 MeV neutrons

Naqvi, A.A. and Khiari, F.Z. and Coban, A. and Aksoy, A. and Al-Jalal, M.A. and Raashid, M. (1992) Measurement of light output of NE213 and NE102A detectors for2.7-14.5 MeV neutrons. Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE, 1.

[img]
Preview
PDF
18Kb
[img]Microsoft Word
26Kb

Abstract

The light output of 125-mm-diameter NE213 and NE102A detectors has been measured for neutron energies ranging from 2.7 to 14.5 MeV. For neutron energies below 6.14 MeV, measurements were carried out using the neutron time-of-flight spectrum from an Am-Be neutron source, while for proton energies above 6.14 MeV, measurements were carried out using neutrons produced from the T(d,n) reaction. For the NE102A detector the measured light output is in good agreement with the data of R.A. Cecil et al., (1979) but for the NE213 detector the light output is 2-15% lower than that for a similar detector. The NE213 detector light output agrees with the data of V. Verbinski et al. (1968)



Item Type:Article
Date:October 1992
Date Type:Publication
Subjects:Computer
Divisions:College Of Sciences > Physics Dept
Creators:Naqvi, A.A. and Khiari, F.Z. and Coban, A. and Aksoy, A. and Al-Jalal, M.A. and Raashid, M.
ID Code:14689
Deposited By:KFUPM ePrints Admin
Deposited On:24 Jun 2008 16:45
Last Modified:12 Apr 2011 13:17

Repository Staff Only: item control page